Hello, welcome to visitJinan Hengsi Shanda Instrument Co., Ltd.
Some of you are interested in:

JBW-300HD ultra-low temperature impact test machine

JBW-300HD ultra-low temperature impact test machine

JBW-300HD ultra-low temperature impact test machine

Product Classification: Impact testing machine
Product Overview:The machine consists of an instrumented impact testing machine, a liquid nitrogen cooling device, and an automatic feeding positioning device, which is used to accurately test and analyze the impact resistance of materials at low temperatures. The low-temperature device adopts liquid nitrogen refrigeration, with a cooling temperature up to -190°, fast cooling speed, automatic temperature control,

JBW-300HD ultra-low temperature impact test machine


Product Description:
The machine consists of an instrumented impact testing machine, a liquid nitrogen cooling device, and an automatic feeding positioning device, which is used to accurately test and analyze the impact resistance of materials at low temperatures.
The low-temperature device adopts liquid nitrogen refrigeration, with a cooling temperature up to -190°, fast cooling speed, automatic temperature control, automatic alarm prompt, and compressor are original imported brand machines.
The sample box can hold 20 samples at a time.
The feeding and positioning is carried out by pneumatic or electric, with fast speed and accurate positioning.
Technical parameters:
Specifications and models
JBW-300HY
JBW-450HY
JBW-750HY
Impact energy
300J
450J
750J
Dial scale range and indexing value
0-300J The index value of each small grid is 2J
0-450J The index value of each small grid is 3J
0-750J each small grid index value 5J
Distance from the axis of the swing shaft to the center of the strike
750mm
750mm
750mm
Shock speed
5.24m/s
5.24m/s
5.24m/s
Sample support span
40 +0.2mm
Arc radius at the end of the sample holder
R1.25mm
Impact blade arc radius
R2.25, R8
Impact tool angle
30°±1°
Impact knife thickness
16 mm
Sample specifications
10×10×55 mm
Force sensor accuracy
≤±1%FS
Angular displacement sensor resolution
0.1°
(The above parameters are for reference only, and the detailed parameters are mainly real equipment)