Operation method of microcomputer screen display hydraulic universal test machine
Release time:2018-11-23 source:Jinan Hengsi Shanda Instrument Co., Ltd. Browse:
Whether the test results are correct, in addition to requiring the universal testing machine itself to achieve the specified accuracy, the operator of the universal testing machine must also be familiar with the operating methods of the test machine, various material testing methods, and knowledge about the properties of the material.Operation method of microcomputer screen display hydraulic universal test machineThe description is as follows:
First turn on the computer, enter the control software, then turn on the main power of the hydraulic universal testing machine, press the "Oil Pump" button on the control panel to start the oil pump motor. Since there is a small amount of air in the pipeline, noise will be generated when the oil flows in the pipeline. The method to eliminate it is to open the oil supply valve to raise the piston for a distance, then close the oil supply valve and open the oil return valve, and then lower the piston. After circulating for a period of time, the air can be discharged before the formal test can be carried out.
The oil feed valve can be opened larger when lifting the test bench, so that the test bench can rise at a fast speed and reduce the auxiliary time of the test. When the sample is clamped and loaded, you should pay attention to the operation. It must be adjusted according to the loading speed of the sample specifications. It should not rise too fast, causing the sample to be impacted, nor should it be closed without reason, causing the load on the sample to suddenly drop, thus affecting the accuracy of the test data. Unless it is done with special specifications of yield points or other special tests, the load needs to be repeatedly increased and decreased, and smooth operation is also required.
When the oil return valve is loaded, it must be closed tightly and no oil return is allowed. After the sample breaks, the oil supply valve should be closed first, and then slowly open the oil return valve to remove the load and let the tester piston fall back to its original position, so that the oil returns to the oil tank. Note: Do not overtighten the handwheel of the oil supply valve to avoid damaging the oil needle. The handwheel of the oil return valve must be tightened, because the oil needle has a large obtuse angle, so it is not easy to damage. (The article was edited and compiled by Hengsi Shanda http://www.hssdtest.com/. If you want to reprint, please keep the source of the connection!)
Recommended productsPRODUCTS