Company News
Operation and maintenance of tensile testing machine
Release time:2018-11-23 source:Jinan Hengsi Shanda Instrument Co., Ltd. Browse:
1. The power supply of the test machine host is not illuminated and cannot move up and down. The solution is to check whether the power line connected to the test machine is properly connected;
Check whether the emergency stop switch is in a screw-up state; check whether the power supply voltage connected to the test machine is normal;
Check whether the safety on the machine socket is blown, please remove the spare fuse and install it.
2. During the daily operation of the force test machine, many problems will occur. Common ones include the message box that appears after the computer software is online.
The solution is to check whether the communication line between the computer and the test machine is off;
Check whether the online selection sensor is selected correctly;
Check whether the sensor has been hit during recent tests or operating the keyboard; check whether the software's calibration or calibration function was used before a problem occurs;
Check whether the calibration values, calibration values, or other information in hardware parameters have been manually changed.
3. The power supply of the test machine host is powered but the equipment cannot move up and down.
The solution is to check whether the device cannot move after 15S (time), because the host needs to be self-checked, which takes about 15S time;
Check whether the upper and lower limits are in the right position and have a certain operating space; check whether the power supply voltage connected to the test machine is normal.
4. The main unit of the test machine adopts a double-screw mid-beam transmission mechanism, which is placed under the hydraulic cylinder, which is convenient to install the sample, has good stability and beautiful shape. The fuel tank adopts a fully enclosed structure, which can effectively prevent dust and other debris from entering the hydraulic system, thereby improving the reliability of the hydraulic system.
The digital display universal test model adopts an LCD screen display measurement system, which can select the test method through the panel buttons and set multiple test parameters.
Recommended productsPRODUCTS