Hello, welcome to visitJinan Hengsi Shanda Instrument Co., Ltd.
Some of you are interested in:
Current location:front page >> News >> Company News

Company News

What are the applicable standards for tension detectors

Release time:2018-11-23 source:Jinan Hengsi Shanda Instrument Co., Ltd. Browse:


Tension testerIt has a wide range of application. Its allowable errors in the force measurement system, extensometer level, and all allowable errors in the extensometer, and all-round test machine instrument standards will be taken to you by Jinan Hengsi Shanda Instrument Co., Ltd. to learn more details.
Table 1: Allowable error of force measurement system
Precision level
Allowed value %
Relative error of display value
q
Relative error of display value repeatability
b
Relative error of the value return journey
v
Relative error of zero point
F0
Relative resolution
a
0.5
±0.5
0.5
±0.75
±0.25
0.25
1
±1.0
1.0
±1.5
±0.5
0.5
Table 2: Allowed coaxiality values
Precision level
Coaxiality allowable value %
0.5
12
1
15
Table 3:Tension testerRelative error of beam movement speed
Precision level
0.5
1
Relative error of crossbeam movement speed %
±0.5
±1
Table 4: Allowed drift value of force measurement system
Precision level
0.5
1
Zero point drift %
±0.5
±1
Table 5: Allowable errors of extensometer levels and extensometers
Extension meter
Precision level
Allowed values
Relative error of gauge distance
Qle
%
Resolution a
Display value error a
Relative error of the value return journey
μ
%
relatively
re/li
%
re
µm
Relative error
qe
%
error
li-lt
µm
0.5
±0.5
0.25
0.5
±0.5
±1.5
±0.75
1
±1.0
0.50
1.0
±1.0
±3.0
±1.5
More AboutMaterial testing machineAll the materials are available in Jinan Hengsi Shanda Instrument Co., Ltd.