Factors affecting the dial pointer of the hydraulic universal test machine

Release time:2018-11-23 source:Jinan Hengsi Shanda Instrument Co., Ltd. Browse:


Hydraulic universal testing machineThe dial pointer has poor sensitivity, and it stops in the middle after unloading, or the zero position often changes. The causes of this phenomenon are multifaceted.
1. If the buffer oil return condition is poor, it should be adjusted or cleaned.
2. The pulleys and tracks on the tooth rods are too dusty and rust. At this time, you need to remove and clean it, and add a little watch oil.
3. The pointer rotation shaft is dirty and should be cleaned.
4. If there is contact between the tooth rod tablet and the tooth rod, it should be loosened.
5. The belt on the force-testing piston is falling off or too loose (makes the piston unable to operate at a uniform speed).
6. If the friction force of the force-testing piston increases or gets stuck in the oil cylinder, use chromium oxide grinding paste to grind until it is normal.
7. There are obstacles during the lifting process of the pendulum, or the pendulum shaft is too dirty or rusted, causing the pointer to return to zero and change greatly.
MoreTesting machineAll information is available at Jinan Hengsi Shanda Instrument Co., Ltd.